Answers are grounded in this paper and cite their sources.
Technical guidelines for deploying the nautilus endovascular device in treating wide-necked aneurysms, highlighting its effectiveness.
The Nautilus intrasaccular system (EndoStream Medical, Israel) is a spiral-shaped neck-bridging endovascular device designed to support coiling of intracranial aneurysms [1–3]. It is deployed into the aneurysm sac through a 0.0165” or 0.017” microcatheter. Coils are delivered via a second microcatheter passed through the struts of the device. The system is available in 4–7-mm sizes and is entirely radio-opaque. It is mechanically detached and fully resheathable. Constructed from a nitinol wire with a platinum-based coil sleeve, it lacks a proximal marker. It has the lowest profile of current intrasaccular devices but a limited size range. Antiplatelet therapy is not mandatory. Limited published series and post-market data suggest high occlusion rates with low complication rates. The device may be particularly suitable for ruptured wide-necked aneurysms. This technical video discusses the device with the aid of two illustrative cases.
No takes yet. Share an insight, caveat, or question.
Lynch et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: