This system measures grip strength in various postures, suggesting its use for rehabilitation monitoring.
Handgrip pressure is a crucial biomechanical indicator, reflecting limb strength and vertebral bone fracture risk. This study introduces a non‐invasive Smart Grip strength measurement system utilizing a Fiber Bragg Grating (FBG) sensor. The device quantifies grip pressure exerted by volunteers in various postures through FBG wavelength shift analysis. Demonstrating a sensitivity of 1.32 pm/με, the system measures grip pressures ranging from 6 to 131 kPa across different hand positions. Tested on 20 volunteers aged 18–25, the maximum recorded grip pressure is 131.9 kPa in the arm‐dangled position. The FBG‐based Smart Grip system provides a precise and efficient method for handgrip strength assessment, valuable for clinical rehabilitation. The novelty of this work lies in the integration of a single FBG‐based Smart Grip system for dual applications such as (1) real‐time posture‐dependent grip strength monitoring and (2) progressive assessment of bone fracture healing.
No takes yet. Share an insight, caveat, or question.
Upadhyaya et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: