Comparison of inertial measurement units and camera systems for spatio-temporal parameters in wheelchair users indicates clinical applicability.
Background: Manual wheelchair propulsion is often associated with pain in the upper extremities. Recording spatio-temporal parameters can optimize movement patterns and prevent injuries. This study compares a marker-based camera system with inertial measurement units to validate their use in wheelchair propulsion on a test stand. Methods: Spatio-temporal parameters of 27 manual wheelchair users propelling at three self-selected speeds (slow, normal, fast) were simultaneously recorded using a marker-based camera system and inertial measurement units, and subsequently compared between both systems. Results: A high correlation was observed among all spatio-temporal parameters (ρ > 0.992). The biases for the start time of hand contact with the pushrim (−0.02 ± 0.02 s), hand release from the pushrim (−0.02 ± 0.01 s), and push length (−0.45 ± 21.45 ms) were slightly overestimated, while recovery length (0.54 ± 21.02 ms), cycle speed (2.37 ± 2.67°/s), and push angle (1.75 ± 4.14°) were slightly underestimated. No bias was found for propulsion frequency. Conclusions: The spatio-temporal parameters recorded using inertial measurement units are suitable for the evaluation of manual wheelchair propulsion and can be used in a clinical context. The low acquisition costs and simple installation process may increase the use of inertial measurement units in the future.
No takes yet. Share an insight, caveat, or question.
Karner et al. (2025) studied this question.