Experimental validation of line laser triangulation for accurate gear evaluation, indicating enhanced precision.
Line laser measurement, as a typical method of laser triangulation, makes the acquisition of 3D tooth-surface data more accurate, efficient, and informative. Thus, a line laser 3D measurement model of gears is established, and a specialized polyhedral artifact with specific geometric features is invented to determine the pose parameters of the line laser sensor in measuring space. Based on this, a single-spindle gear-measuring instrument is developed and a series of experimental studies are conducted for gears with different module and flank directions in this instrument, including profile deviation, helix deviation, pitch deviation, topological deviation, etc. A comparative experiment with traditional contact measurement methods validates the correctness of the methods mentioned in this paper for the accurate evaluation of tested gears. In further research, the mining and utilization of big data obtained from the line laser 3D measurement of gears will be an important topic.