Analysis demonstrates reduced uncertainty in yield and ion-temperature measurements in DT implosions, highlighting improved accuracy in hotspot flow assessments.
A new neutron time-of-flight (nTOF) detector for yield and ion-temperature measurements in deuterium-tritium (DT) implosions was designed, installed, and calibrated on the OMEGA Laser System. This detector was installed at θ = 46.8° from the vertical and φ = 84.9°, previously a region without other detectors, in order to decrease the uncertainty of low-mode asymmetry measurements by six other nTOF detectors. The new nTOF detector uses the "PMT only" concept [Glebov et al., Rev. Sci. Instrum. 92, 013509 (2021)] and consists of a Photek photomultiplier tube (PMT)-110 in a thin aluminum housing. There is no lead shielding in front of the PMT to allow measurements of the x-ray instrument response function of the detector. The detector is installed in the OMEGA H2-port line of sight at 9.3 m from the target chamber center in the OMEGA Target Bay and is called the H2 nTOF detector. The H2 nTOF measures DT yields in the 1 × 1013 to 3 × 1014 range and ion temperatures from 2 to 12 keV. H2 nTOF reduced the hotspot flow uncertainty in the z-direction from ±10 to ±8 km/s, and the uncertainty in the anisotropic flow axis from σθ = ±26° to ±13° and σφ = ±25° to ±21°. The design details, performance, and calibration results of the H2 nTOF detector in DT implosions on OMEGA will be presented.