Answers are grounded in this paper and cite their sources.
Novel IR-EPI scheme improves T1 mapping accuracy in cerebral imaging, suggesting rapid acquisition benefits.
Motivation: The inversion-recovery method for T1 mapping typically requires a relatively long acquisition time which can be significantly shortened using EPI techniques. However, substantial geometric distortions in EPI often hinder accurate T1 quantification. Goal(s): We propose a novel IR-EPI scheme that simultaneously acquires EPI data with the reversed phase-encoding direction in a single IR measurement. Approach: Geometric distortions of IR-EPI data were corrected using the blip-up/blip-down approach. Its performance was evaluated in direct comparison to that of original IR-EPI. Results: The proposed scheme enables effective distortion correction for IR-EPI data, enhancing T1 mapping accuracy while providing rapid, whole-cerebrum acquisition at 7T. Impact: The proposed IR-EPI scheme enables effective geometric distortion correction in IR-EPI data, with only a negligible increase in scan time. Consequently, the greater structural similarity to co-registered MP2RAGE significantly enhances T1 mapping accuracy, while providing rapid, whole-cerebrum acquisition at 7T.
No takes yet. Share an insight, caveat, or question.
Yun et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: