Answers are grounded in this paper and cite their sources.
Observational analysis compared intraoperative and conventional MRS coils in brain regions, suggesting ioMRS enhances cancer treatment tracking.
Motivation: Intraoperative magnetic resonance spectroscopy (ioMRS) has rarely been applied, although it offers high potential. It could be introduced to MR-guided laser-induced thermal therapy (MRgLITT) for improving cancer treatment. Goal(s): The aim was to compare two MRS sequences measured with a conventional and an intraoperative (io) coil at different brain regions. Approach: 10 healthy subjects were measured at 3T with a conventional and an io coil using a PRESS and sLASER sequence in the cingulate gyrus and amygdala. Results: Better spectral quality was gained with the conventional coil in regions closer to the coil. No differences in concentrations were found between the coils. Impact: Intraoperative MR spectroscopy (ioMRS) is feasible and has been performed for the first time with an sLASER sequence. Future implementation in MR-guided laser-induced thermal therapy (MRgLITT) allows tracking of metabolic changes due to LITT.
No takes yet. Share an insight, caveat, or question.
Pfleger et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: