This research demonstrates enhanced ultrasound localization in MR-ARFI, suggesting a reduction in energy deposition for improved safety.
Motivation: Magnetic resonance acoustic radiation force imaging (MR-ARFI) is a promising tool for ultrasound localization in MR-guided ultrasound applications. To ensure image quality, repetitive ultrasound sonication is required and may cause inevitable neurological effects and energy deposition. Goal(s): We aim to develop a reconstruction algorithm for highly under-sampled MR-ARFI. Approach: The reconstruction is formulated by a constrained minimization problem regulated by the spatial sparsity of displacement image and data consistencies between ARFI images with opposite displacement encodings. Results: Experimental results of retrospective and prospective under-sampling show that the proposed method is well-performed in terms of reconstruction error at acceleration rate as high as 10. Impact: The proposed method can be exploited to achieve fast MR-ARFI at high under-sampling rate to reduce the ultrasound deposition and shorten the scanning time, which is crucial for the safety concern of ARFI application in human study.
No takes yet. Share an insight, caveat, or question.
Yuan et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: