Innovative coil improves SNR in breast interventions with low-field MRI, highlighting mobile medical options.
Motivation: As breast cancer becomes the most common malignant tumor in women, MRI-guided breast interventions have demonstrated their practical value in clinical. Goal(s): Low-field MRI have emerged as a new option for breast intervention platforms due to their high openness, low cost and mobility. Approach: We designed an adjustable surface coil for low-field MRI-guided breast interventions. MR images and SNR maps obtained with the surface coil and volume coil were compared using phantoms and in vivo human breast experiments. Results: The SNR of images from the silicone phantom and human breast exceed 80 dB, approximately 15% ~ 30% higher than that of the volumetric coil. Impact: This adjustable surface coil is designed to work with the open low-field MRI system, providing high imaging SNR for real-time guidance during breast interventional surgeries and expanding access to precise medical care in remote areas through the mobile MRI platform.
No takes yet. Share an insight, caveat, or question.
Liu et al. (2025) studied this question.