Automatic velocity encoding calibration optimizes flow measurement in healthy volunteers, simplifying processes.
Motivation: The optimal velocity encoding limit (VENC) in phase contrast MRI is subject-specific and is currently manually set by the technologist in a time-consuming process. Goal(s): We propose an inline automatic velocity encoding calibration for 2D phase-contrast flow MRI, which automatically computes and sets the optimal VENC. Approach: We designed a calibration module to automatically estimate the optimal VENC inline and set it in the subsequent flow measurement. Sixteen healthy volunteers were imaged at 0.55T. Results: Deployed inline, the calibration module calculated correctly the optimal VENC, mean of 97% of the maximum velocity in the ascending aorta, and automatically applied for subsequent flow sequence. Impact: Inline automatic velocity encoding calibration ensures the optimal precision in flow measurements and simplifies the acquisition workflow.
No takes yet. Share an insight, caveat, or question.
Daudé et al. (2025) studied this question.