Why the study?
The study investigated whether velocity encoding adjustment is necessary for phase-contrast flow volume magnetic resonance measurements when using modern scanners with effective background velocity offset corrections.
Are reliable phase-contrast flow volume magnetic resonance measurements feasible without adjustment of the velocity encoding parameter in patients with aortic regurgitation?
Are reliable phase-contrast flow volume magnetic resonance measurements feasible without adjustment of the velocity encoding parameter in patients with aortic regurgitation?
Phase-contrast flow volume measurements can be reliably performed without adjusting the VENC parameter on modern MR scanners, potentially reducing scan time.
Answers are grounded in this paper and cite their sources.
Supports omitting VENC adjustment in modern MR scanners; leaves open validation in larger, diverse cohorts before routine use.
Purpose: To show that adjustment of velocity encoding (VENC) for phase-contrast (PC) flow volume measurements is not necessary in modern MR scanners with effective background velocity offset corrections. Approach: The independence on VENC was demonstrated theoretically, but also experimentally on dedicated phantoms and on patients with chronic aortic regurgitation (n = 17) and one healthy volunteer. All PC measurements were performed using a modern MR scanner, where the pre-emphasis circuit but also a subsequent post-processing filter were used for effective correction of background velocity offset errors. Results: The VENC level strongly affected the velocity noise level in the PC images and, hence, the estimated peak flow velocity. However, neither the regurgitant blood flow volume nor the mean flow velocity displayed any clinically relevant dependency on the VENC level. Also, the background velocity offset was shown to be close to zero (<0.6 cm / s) for a VENC range of 150 to 500 cm / s, adding no significant errors to the PC flow volume measurement. Conclusions: Our study shows that reliable PC flow volume measurements are feasible without adjustment of the VENC parameter. Without the need for VENC adjustments, the scan time can be reduced for the benefit of the patient.
No takes yet. Share an insight, caveat, or question.
Lagerstrand et al. (2020) studied this question.