Evaluation reports the impact of eddy currents and mechanical oscillations on phase errors in phase contrast MRI, suggesting corrections depend on velocity encoding gradients.
Motivation: Background phase errors in PC-MRI include eddy currents and mechanical oscillations. Relative contributions over a wide range of velocity encoding gradients (VENCs) remain underexplored. Goal(s): To evaluate background phase errors across clinically relevant VENCs, characterizing the impact of eddy currents and mechanical oscillations. Approach: Waveforms with a range of VENCs were generated, and the gradient impulse response function was used to predict phase errors as a sum of eddy current and mechanical oscillation effects. Phantom experiments across different VENCs were conducted. Results: Eddy current effects dominated lower VENCs. Between approximately 52 and 92 cm/s the phase error was largely oscillatory. Impact: The necessary correction for background phase errors depends on the VENCs, which explains why errors are protocol dependent.
No takes yet. Share an insight, caveat, or question.
Thalén et al. (2025) studied this question.