This research demonstrates diffusion frequency-dependence in an anisotropic phantom, indicating OGSE is effective for microstructural imaging on high-gradient systems.
Motivation: Diffusion frequency-dependence offers a direct probe to tissue microstructure at diffusion length scale. Oscillating-gradient-spin-echo (OGSE) sequence provides very short effective diffusion time and length for evaluating fine structures but poses challenges in simultaneously achieving high diffusion-weighting and high frequency on human scanners. Goal(s): To validate the potential of OGSE sequences on the high-gradient-performance Connectome 2.0 scanner. Approach: We measured diffusivity frequency-dependence (30-100Hz) in an anisotropic solid-fiber phantom on Connectome 2.0 and estimated surface-to-volume ratio and fiber radius, whose value was corrected for non-ideal cosine-OGSE waveform. Results: Estimated fiber radius matched the ground-truth (error<5%), demonstrating the potential of in vivo microstructural imaging using OGSE. Impact: We performed diffusion frequency-dependence using oscillating-gradient-spin-echo (OGSE) sequence (≤100Hz) and accurately estimated fiber radius in an anisotropic fiber phantom on the Connectome 2.0 human scanner, demonstrating the potential of in vivo microstructural imaging using OGSE.
No takes yet. Share an insight, caveat, or question.
Zheng et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: