Answers are grounded in this paper and cite their sources.
Novel TGSE-PROPELLER diffusion-weighted imaging achieves distortion-free high resolution, suggesting improved prostate imaging protocols.
Motivation: Clinical single shot EPI-DWI suffers from low resolution and geometric distortion. Goal(s): To achieve distortion-free, high-resolution prostate DWI within the same or shorter scan time as the clinical routine. Approach: We developed a novel Turbo Gradient Spin Echo (TGSE) PROPELLER DWI acquisition scheme for prostate imaging, incorporating a self-supervised model for denoising. Blades corresponding to the same b-values are combined to produce b50 and b1000 images for ADC fitting. Results: DWI images generated with TGSE-PROPELLER-DWI are distortion-free, even in cases with metal implants or strong air-tissue induced B0 inhomogeneity. Higher-than-clinical resolution is achievable with the same scan time. Impact: TGSE-PROPELLER-DWI with a novel acquisition scheme provides distortion-free, high-resolution prostate DWI within the same scan time as the clinical protocol, with potential for further scan time reduction.
No takes yet. Share an insight, caveat, or question.
Chen et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: