Evaluation reveals no change in quantitative diffusion tensor imaging metrics with DL-denoising, suggesting stability in muscle data.
Motivation: Deep learning (DL) denoising reconstruction reduces noise without added scan times. However, the effects of this denoising on quantitative diffusion tensor imaging (DTI) metrics have yet to be evaluated. Goal(s): Determine the effects of DL-denoising, across a range of signal averages (NEX), on quantitative DTI metrics in lower leg muscles. Approach: 6 Subjects obtained a series of DTI acquisitions. Quantitative biometrics of muscle compartments were compared with and without DL-denoising. Results: Biometric accuracy for each NEX was similar with and without DL-denoising. All coefficients of variation fell below 5%. Concordance correlation coefficients were above 0.87. Impact: The influence of DL-denoising on quantitative diffusion tensor imaging (DTI) metrics is unknown. We found that DL-denoising in lower leg muscles does not affect quantitative DTI measures but also does not improve accuracy at lower signal averages.
No takes yet. Share an insight, caveat, or question.
George et al. (2025) studied this question.