The DES-QUTE sequence quantifies T2* and diffusion coefficients efficiently, highlighting its utility in MRI imaging.
Motivation: T2* relaxation time is typically assessed using a multi-echo GRE sequence. However, the accurate quantification of slow-relaxing water pools, such as CSF, necessitates an extended scan time in comparison to that required for the quantification of brain tissue. Goal(s): We present a new sequence (DES-QUTE), which is capable of acquiring three bunches of echoes corresponding to short, intermediate, and long TEs without a significant increase in the scan time. Approach: The proposed sequence integrates QUTE with an improved echo-shifting technique. Results: The DES-QUTE demonstrated accurate and precise T2* quantification at ultra-high field MRI, while providing a potential for simultaneous quantification of diffusion coefficients. Impact: The proposed sequence offers an opportunity to sample different ranges of TE images and quantify both T2* and D information without a significant increase in scan time. This method is expected to make significant contributions to the study of neurofluids.
No takes yet. Share an insight, caveat, or question.
Shin et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: