This approach reduces scan time by 3x through compressed sensing with decoupled solenoidal array coils.
Motivation: Accelerate wrist imaging on low field point of care MRI. Goal(s): Introducing a solenoidal phased array coil to reduce scan time with compressed sensing. Approach: Using electrical and geometrical decoupling methods to create three inherently decoupled solenoidal phased array and accelerating scan by using compressed sensing. Results: Inter-element coupling, and S-parameters for each of the three individual Rx coils and the transmit coil under loaded conditions were lower than -25 dB. A 3x compressed sensing acceleration produced results comparable to fully sampled k-space. Impact: This method of decoupling can make the application of solenoidal phase array possible for wrist imaging accelerate by factor 3 which is very important in point of care MRI systems.
No takes yet. Share an insight, caveat, or question.
Parsa et al. (2025) studied this question.