Answers are grounded in this paper and cite their sources.
This approach reduces scan time by 3x through compressed sensing with decoupled solenoidal array coils.
Motivation: Accelerate wrist imaging on low field point of care MRI. Goal(s): Introducing a solenoidal phased array coil to reduce scan time with compressed sensing. Approach: Using electrical and geometrical decoupling methods to create three inherently decoupled solenoidal phased array and accelerating scan by using compressed sensing. Results: Inter-element coupling, and S-parameters for each of the three individual Rx coils and the transmit coil under loaded conditions were lower than -25 dB. A 3x compressed sensing acceleration produced results comparable to fully sampled k-space. Impact: This method of decoupling can make the application of solenoidal phase array possible for wrist imaging accelerate by factor 3 which is very important in point of care MRI systems.
No takes yet. Share an insight, caveat, or question.
Parsa et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: