Answers are grounded in this paper and cite their sources.
In vivo evaluation shows dual-saturation combined with multiband enhances myocardial coverage, suggesting improved diagnostics.
Motivation: Quantitative myocardial perfusion is limited to capturing 3 slices per heartbeat. This limitation highlights the need to increase the number of slices, aiming for full left ventricular coverage with absolute myocardial blood flow quantification. Goal(s): To integrate quantitative perfusion imaging with simultaneous multi-slice (or multiband) excitation. Approach: Multiband was integrated into an existing dual-saturation perfusion sequence. In-vivo experiments in healthy sedated pigs were conducted to evaluate the reliability of the technique under rest conditions. Results: Combining dual-saturation sequences with multiband is feasible. The number of slices acquired with multiband was duplicated relative to gold standard. Myocardial Blood flow values were comparable and reproducible. Impact: Combining quantitative perfusion imaging with multiband techniques expands myocardial coverage while ensuring precise blood flow quantitative measurements. This integration represents a significant advancement in potentially enhancing diagnostic capabilities in cardiovascular imaging.
No takes yet. Share an insight, caveat, or question.
Ferrazzi et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: