Innovative design improves liquid dispensing in Lab-on-Chip systems for Point-of-Care environments, suggesting enhanced diagnostic precision.
This paper presents an innovative and low-cost approach to the dispensing of multiple liquids on a microfluidic chip with the aim of dispensing liquids in a controlled sequence. The project focused on the design and development of a microfluidic liquid dispensing system that is an integral part of the Lab-on-Chip (LOC). Liquids are often dispensed into LOCs through blisters, syringes, or electric microfluidic pumps, but these can be impractical for Point-of-Care (POC) settings, especially in remote areas. Additionally, incorrect volumes of biochemical reagents and the introduction of reagents outside the sequence can distort the results of the diagnosis. The process undertaken involved designing and 3D printing prototypes of the dispensing system, along with laser cutting and manufacturing the Polymethyl Methacrylate (PMMA) LOC devices intended for receiving the liquids. The proposed novel low-cost dispensing system uses manually operated actuators and cams to disperse metered fluids sequentially to minimise end-user errors at POC settings.
No takes yet. Share an insight, caveat, or question.
Kakaza et al. (2025) studied this question.