Development of a self-calibrating radiation source improves spectral irradiance measurement accuracy in lunar observation, indicating high stability.
To enhance the field calibration capability of ground-based lunar observation instruments for long-term continuous monitoring and to optimize the stability and traceability of lunar observation data, this manuscript presents the development of a SI traceable Portable Self-calibrating Absolute Radiation Source (PSARS) based on an electrical substitute radiometer. A self-calibrating radiation transfer model has been established. The system features a “+” structure layout centered around an integrating sphere, which ensures uniformity of the light source while improving system integration. Preliminary performance testing results indicate that PSARS achieves excellent radiative planar uniformity and angular uniformity within the targeted area, both exceeding 99%. During the self-calibration cycle of PSARS, the detector demonstrates high measurement stability for the built-in light source. Ultimately, through comparative validation and uncertainty assessment, the self-calibration accuracy of spectral irradiance for PSARS in the 400–1000 nm wavelength range is better than 2%, meeting the demands for high-frequency, high-stability, and high-precision real-time on-site radiometric calibration under ground-based lunar observation field test conditions. This provides technical support for the construction of high-precision lunar models and the widespread application of lunar calibration technologies.