Analytical model relates motion artifacts to pulse signal distortion in tactile sensors, highlighting design impacts.
This paper, the second of two parts, presents an analytical model of motion artifacts (MA) in measured pulse signals by a tactile sensor, which contains a deformable microstructure sitting on a substrate. While the tissue-contact-sensor (TCS) stack and the sensor are both treated as a 1DOF (degree-of-freedom) system, tissue-sensor contact joins their mass together to form a 1DOF system with springs and dampers on both sides. MA on the sensor substrate causes baseline drift and time-varying system parameters (TVSP) of the TCS stack simultaneously. An analytical model is developed to mathematically relate baseline drift and TVSP to a measured pulse signal. The numerical calculation is conducted in MATLAB. Baseline drift in a measured pulse signal is much lower than the actual MA in its measurement. As compared to baseline drift, TVSP generates relatively abrupt, small distortion (e.g., 0.2% variation in heart rate and <5% change in pulse amplitude), but it rides on each harmonic of the true pulse signal. Sensor design alters both the deviation of the amplitude and waveform of a measured pulse signal from the true pulse signal and the influence of MA on it.