A 30-month field evaluation of low-cost CO2 sensors using a reference instrument

View Full Paper
QCQixiang CaiNZNing ZengXYXiaoyu Yang

Key Points

  • The environmental correction system reduced root mean square errors from 5.9 ppm to 1.6 ppm, improving sensor accuracy.
  • Long-term drift produced biases up to 27.9 ppm over two years, affecting the overall performance of low-cost sensors.
  • Recommendations include calibrating low-cost sensors every three to six months to maintain accuracy within 5 ppm.
  • SENSE-IAP sensors demonstrated capability for long-term operation without extensive recalibration, saving time and costs.

Abstract

Abstract. Low-cost, medium-precision sensor (LCS) networks have emerged as a promising approach for CO2 monitoring under complex urban emission conditions. However, the field performance of these LCSs faces significant challenges from environmental factors (e.g., temperature, humidity) and long-term drifts caused by sensor degradation (e.g., light source aging). In this study, we performed 30 months of co-located observations using LCS units (named SENSE-IAP) alongside a Picarro reference analyzer to evaluate long-term field performance, which is essential for the correction and validation of mid–low-cost CO2 observation networks. The environmental correction system we developed effectively corrected the impact of daily environmental changes, which reduced the root mean square errors (RMSEs) from 5.9 ± 1.2 to 1.6 ± 0.5 ppm for SENSE-IAP. The corrections remained robust against seasonal environmental changes, maintaining daily RMSE typically within 1–3 ppm throughout the 30 months of observation. Long-term drifts, commonly occurred in LCS, produced biases of up to 27.9 ppm over 2 years. Furthermore, seasonal drift cycle contributed up to 25 ppm RMSE after 6 months of deployment. Although the environmental correction system could not address these errors, linear interpolation effectively calibrated the long-term drift, reducing the 30-month RMSE to 2.4 ± 0.2 ppm. To improve the accuracy of high-density CO2 networks using such LCSs, we recommend maintaining a calibration frequency preferably within 3 months and not exceeding 6 months, with optimal calibration performed during both winter and summer to ensure accuracy within 5 ppm. These findings indicate that SENSE-IAP instruments can operate long-term without requiring a return to the laboratory for re-calibration or frequent field standard gas calibration, thereby substantially reducing time, labor, and financial costs.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Cai et al. (2025) studied this question.

synapsesocial.com/papers/68dc26188a7d58c25ebb283ahttps://doi.org/10.5194/amt-18-4871-2025