Investigating ionization current and optical readout methods for RF electric field sensitivity in cesium vapor cells.
We investigate a technique for detecting radio-frequency (RF) electric fields in a Cesium (Cs) vapor cell at room temperature by collecting charge from ionized Rydberg atoms and compare its performance with the established method of electromagnetically induced transparency (EIT). By applying a known RF field, we measure the response from both the electrical (ionization current-based) and optical (EIT-based) readouts. The ionization current-based method yields a sensitivity of 22~$μ$Vm⁻¹Hz-1/2, while the EIT-based method achieves 3.7~$μ$Vm⁻¹Hz-1/2. The sensitivity of the ionization current-based method is limited by thermal noise arising from a 2.2~k$Ω$ resistance between the collection electrodes, attributed to a thin Cs film on the inner surfaces of the vapor cell. Controlling or eliminating the Cs layer can significantly improve the sensitivity of this ionization approach.
No takes yet. Share an insight, caveat, or question.
Shylla et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: