Answers are grounded in this paper and cite their sources.
This study introduces a new calibration technique that improves measurement accuracy in infrared thermography, suggesting better fever screening outcomes.
Infrared thermography for human skin temperature measurement, when calibrated with standard blackbodies, suffers from errors due to the mismatch in emissivity between a blackbody and human skin. This study introduces a novel calibration method utilizing a human skin-like gradient radiation source to enhance measurement accuracy. A custom radiation source with six temperature points and skin-like emissivity was developed. Thermal imagers were calibrated using this source, and their performance was compared against traditional blackbody calibration. The proposed method reduced the calibration error to 0.04 °C, a significant improvement over the 0.15 °C error obtained with blackbody calibration. Calibration with a skin-like radiation source proves superior to the blackbody method, enabling high-accuracy (less than 0.1 °C) human skin temperature measurement for improved fever screening.
No takes yet. Share an insight, caveat, or question.
Li et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: