Atomic magnetometers (AMs) that use alkali vapors, such as rubidium, are among the most sensitive sensors for magnetic field measurement. They commonly use polarization differential detection to mitigate common-mode noise. Nevertheless, traditional differential detection optics, including polarization beam splitters (PBS) and half-wave plates, are typically bulky and large, which restricts further reductions in sensor dimensions. In this study, a combination of liquid crystal polarization grating (LCPG) and liquid crystal quarter-wave plate is used for differential detection in AMs, with magnetic field strength determined by measuring the intensity of two diffracted beams from the LCPG. The experimental findings indicate that the fabricated LCPG exhibits a circularly polarized extinction ratio of 3,656 and achieves an average diffraction efficiency of 99 %. In addition, the differential detection method based on LCPG can achieve an angular resolution of 1.48 × 10