Background Total testosterone immunoassay is a commonly ordered test for the diagnosis of gonadal disorders. However, immunoassays suffer from low specificity and sensitivity, which leads to inaccuracy especially in the low concentration range (< 100 ng/dL). As a result, commercial testosterone immunoassays have difficulties to accurately measure total testosterone levels for women, children, and hypogonadal men. LC-MS/MS has emerged as the method of choice for the accurate quantitation of steroid hormones. We presented here a sensitive laboratory-developed total testosterone assay, which was validated on two LC-MS/MS instruments. Methods An internal standard containing [2,3,4-13C3]-testosterone was added to each sample. The spiked serum sample was loaded on a supported liquid extraction (SLE) cartridge and eluted using methylene chloride. The elute was dried down under nitrogen and reconstituted in sample reconstitution buffer. Chromatographic separation was achieved using a Restek Ultra C18, 50 x 2.1 mm, 3µm analytical column and a total run time of each sample was 5 minutes. The assay was validated on two triple quadrupole mass spectrometers, API 6500 and API 4500. Results were compared with those of a reference laboratory. Results The assay was linear in the range of 1 to 1200 ng/dL for total testosterone and the AMR was set at 2-1200 ng/dL. The interday and intraday imprecision were less than 8%. Split sample comparison to a reference laboratory demonstrated acceptable accuracy with the slope of 1.053 and correlation coefficient (R2) of 0.991. The limit of detection (LOD) was 0.2 ng/dL and the limit of quantitation (LOQ) was 0.5 ng/dL. Results obtained from two instruments agreed well with the slope of 0.996 and the R2 of 0.999. No carryover or interferences was observed for this assay. Conclusions A LC-MS/MS based method was developed and validated for measuring total testosterone levels for female and children. This sensitive and specific assay will improve the diagnosis of hypogonadism and monitoring of testosterone-suppressing treatment.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Cao et al. (2024) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: