This research investigates the utilization of an ionic liquid combination of solidified floating organic drop micro-extraction (IL-SFODME) to augment the concentration of trace amounts of lead, working as a preliminary stage before electrothermal atomic absorption spectrometry (ETAAS) analysis without the use of chelating agents. Key parameters impacting the microextraction efficiency—including pH, the volume of the ionic liquid (1-Hexyl-3-methylimidazolium hexafluorophosphate, HMIMPF6), temperature, extraction time, and stirring speed—were methodically examined to determine optimal conditions. Under detected optimized conditions, an enhancement factor of 71.2 was obtained for a 15 mL sample solution. The calibration curve exhibited linearity within the concentration range of 0.2–2.5 µg/L, with a detection limit (3σ) of 0.054 µg/L and a quantification limit (10σ) of 0.18 µg/L. For seven replicate measurements of 0.5 µg/L lead, the relative standard deviation (RSD) was ±2.30%. This method was effectively implemented to extract and quantify lead in both reference water and different real water samples, showcasing significantly efficient extraction performance.