The PLACID instrument installed at the new Turkish 4m-DAG telescope in 2024 will be the first active stellar coronagraph, offering a unique facility to perform innovative high-contrast imaging observing strategies. It is composed of a Spatial Light Modulator (SLM) in reflection in lieu of the usual fixed focal plane mask to digitally set up any phase mask in real time. However, such hardware works with a linearly polarized wave causing the loss of 50% throughput. To retrieve full capacity, PLACID will then be upgraded with polarization-independent solutions to be first tested on the Swiss Wideband Active Testbed for Coronagraphic High-contrast imaging (SWATCHi) 2.0. This testbed is also employed to investigate coherent differential imaging techniques that relies on the fast modulation of the speckle intensity whose signal is recorded with a C-RED ONE camera. This work presents the optical design and first results of SWATCHi 2.0.
No takes yet. Share an insight, caveat, or question.
Potier et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: