We present the laboratory performance of a 1,704-detector microwave SQUID multiplexed (μmux) Transition-Edge Sensor (TES) array and its accompanying RFSoC-based readout electronics after integration with AliCPT-1: a 72-cm aperture, two-lens refracting CMB polarization telescope, which will observe from the Tibetan Plateau beginning in 2024. The first AliCPT-1 module incorporates two 4-6 GHz μmux readout lines, each with a multiplexing factor of 910, to readout 1,704 TES bolometers and additional dark channels. The module is probed via an RFSoC-based room-temperature electronics system, which combines all tone generation and processing procedures on a single piece of hardware. Here we describe the integration of the module and electronics with the receiver and conduct end-to-end laboratory testing of their performance. Module measurements include spectra, optical efficiency, polarization efficiency, near-field beam maps, and sensitivity to external magnetic fields. We evaluate the behavior of the readout system during optical data collection, including its noise performance.
No takes yet. Share an insight, caveat, or question.
M. O. Withers (2024) studied this question.