This is the authors' abstract. We don't add key points for this paper.
Embedded sensors in implants that communicate with an MRI system can improve patient safety substantially, in particular when parallel transmission is applied to mitigate the RF-heating risk. Calibration procedures and virtual models are presented in this work to support a rigorous safety assessment of this novel approach to implant safety. Experiments show a good correlation (r>0.96) between sensor signal (voltage at the tip) and field-probe measurements (E-field and temperature) for a custom-built reference implant. These results are further supported by simulations that can be extended to investigate more complex implant heating scenarios in human voxel models.