Balanced Steady State Free Precession (bSSFP) has been successfully used to estimate T2 maps (DESPOT2), but bSSFP is prone to off-resonance artifacts. These artifacts affect the reconstruction of the T2 maps. We propose a dictionary-matching reconstruction using the signal model of bSSFP and spoiled gradient echo (SPGR). We tested this approach with simulation, 3D phantom and 3D in-vivo prostate acquisitions. In all of them, the off-resonance banding artifacts are greatly reduced, which is reflected in reduced NRMSE values. This method allows using a faster sequence with high SNR to quantify complicated body regions like the prostate.
No takes yet. Share an insight, caveat, or question.
Coronado et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: