Accuracy and test-retest repeatability of stiffness measurement with MR Elastography: a multi-center phantom study

View Full Paper
EOEfe OzkayaPKPaul KennedyJCJun Chen

Key Points

Key points are not available for this paper at this time.

Abstract

The purpose of our multicenter study was to determine the accuracy and test-retest repeatability of stiffness measured with MRE in phantoms. Three phantoms with known stiffness (2.4, 4.4, and 7.5 kPa) were circulated between 5 different centers. 1.5T or 3T systems from the 3 major vendors were used for scanning with 2D GRE and/or SE-EPI MRE sequences. The reference was based on measurements in the reference center made at the start and end of study. For all 3 phantoms the mean accuracy error for the 4 testing centers was 10.2%, while test-retest repeatability error was 2.9% for all centers.