The dual flip angle (DFA) method is widely used for brain T1 quantification. In addition to the well-known RF field effect, this study investigated the overlooked magnetization transfer (MT) effect for the accuracy of in-vivo T1 mapping. Based on the two-pool Bloch simulation, the DFA derived T1 estimation revealed FA and TR dependence, which was confirmed by the in-vivo brain T1 mapping using the same set of sequence parameters. Longer TR (30 ms) reduced the variation between FA pairs and were closer to the literature values, but at the cost of longer acquisition time.
No takes yet. Share an insight, caveat, or question.
Hu et al. (2024) studied this question.