August 14, 2024Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition0 citations

RF mode switching based Tx coil system for B1+ homogeneity mitigation at 7T: basic concept and preliminary study

View Full Paper
SUShin‐ichi UrayamaMFMasaki FukunagaMCMartijn A. Cloos

Key Points

Key points are not available for this paper at this time.

Abstract

To mitigate B1 inhomogeneity problems in ultra-high-field MRI, we proposed RF-mode switching technique which doesn’t need an expensive parallel transmit (pTx) system. As a preliminary study, we developed an RF-switching circuit (one of the main parts of the system) and validated its function through phantom experiments. The results showed that the switching function works well, although a certain degree of power-loss was observed both at bench (< -1dB) and on the images. We have re-designed a new switching circuit and will continue to the development to demonstrate this technique at a fraction of the cost of a pTx system.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Urayama et al. (2024) studied this question.

synapsesocial.com/papers/68e5c52db6db64358755c03chttps://doi.org/10.58530/2023/4412
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Slice‐selective RF pulses for in vivo B inhomogeneity mitigation at 7 tesla using parallel RF excitation with a 16‐element coil2008 · 164 citations
  2. 2MRI B1 Optimization using Coupled Coil Modal Analysis2025
  3. 3B1+ inhomogeneity mitigation in slice-selective excitation at 7T with a single-channel RF coil augmented by ΔB0 shim array fields2024
  4. 4Design of a local transmit RF-resonator with enhanced local B1+-field for improved abdominal imaging in a single-sided low-field system2025
  5. 5Coaxial Dipole Array With Switching Transmit Sensitivities for Ultrahigh Field MRI2026