Safe access to MRI for patients with active implants may be possible by measuring RF currents in implants in individual patients with low-SAR sequences. We present a simple theoretical model of the B1+-field close to a locally straight wire at any angle to B0. MRI signals fitted by our model closely match magnitude and complex signals from electromagnetic simulations with a wire parallel to B0 or tilted at 45° and magnitude experimental signals with parallel wire. RF currents reconstructed from simulated data match ground-truth values to within 6% and current profiles reconstructed from experimental signals follow simulated currents in shape.
No takes yet. Share an insight, caveat, or question.
Hartmann et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: