Magnetic resonance-guided high-intensity focused ultrasound (MRgHIFU) treatments are limited by motion artifacts introduced into MR thermometry calculations by large sporadic motions. A hybrid principal component analysis and projection onto dipole fields (PCA-PDF) motion compensation algorithm was expanded upon to include predicted motion using image augmentation techniques. When tested on MR images from a gelatin phantom being translated in the coronal plane, the improved PCA-PDF algorithm yielded a temperature standard deviation of 0.4 ± 0.1 °C, which is a 3.7 ± 0.1 °C reduction from an unmodified PCA-PDF approach, indicating that thermometry artifacts induced by motion were significantly reduced.