An electromagnetic design of a 16.0 T MRI magnet using a combination of Nb3Sn and NbTi superconductors has been completed. It consists of a 9.0 T inner Nb3Sn solenoid and a set of outer NbTi coils that produces 7.0 T field, for a combined 16.0 T field. The design revealed very high hoop tensile and axial compressional stresses in the Nb3Sn coil. Alloy-reinforced high-strength Nb3Sn wires were selected for use in this project. For a magnet to operate in a persistent mode, superconducting joints between Nb3Sn wire segments need to be developed. Various joint configurations were attempted with some success.