The localization of EEG sources is one of the fundamental approaches to facilitate the interpretation of EEG data. However, the accuracy of source localization depends on the exact knowledge of the position of the electrodes on the scalp, which currently requires time-consuming and/or expensive approaches. Here, an automatic method is proposed that retrieves the electrode positions by localizing the curvature changes in T1-weighted MRI images caused by electrodes. The results show an average detection sensitivity of ~96.4%, with an average position error of 4.23 mm for all subjects.