The objective of this study was to evaluate a high-resolution deep-learning (DL)-based diffusion-weighted imaging (DWI) sequence for breast magnetic resonance imaging (MRI) in comparison to a standard DWI sequence (DWIStd) at 1.5 T. It is a prospective study of 38 breast cancer patients, who were scanned with DWIStd and DWIDL. Both DWI sequences were scored for image quality, sharpness, artifacts, contrast, noise, and diagnostic confidence with a Likert-scale from 1 (non-diagnostic) to 5 (excellent). The lesion diameter was evaluated on b 800 DWI, apparent diffusion coefficient (ADC), and the second subtraction (SUB) of the contrast-enhanced T1 VIBE. SNR was also calculated. Statistics included correlation analyses and paired t-tests. High-resolution DWIDL offered significantly superior image quality, sharpness, noise, contrast, and diagnostic confidence (each p < 0.02)). Artifacts were significantly higher in DWIDL by one reader (M = 4.62 vs. 4.36 Likert scale, p < 0.01) without affecting the diagnostic confidence. SNR was higher in DWIDL for b 50 and ADC maps (each p = 0.07). Acquisition time was reduced by 22% in DWIDL. The lesion diameters in DWI b 800DL and Std and ADCDL and Std were respectively 6% lower compared to the 2nd SUB. A DL-based diffusion sequence at 1.5 T in breast MRI offers a higher resolution and a faster acquisition, including only minimally more artefacts without affecting the diagnostic confidence.
No takes yet. Share an insight, caveat, or question.
Olthof et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: