Kelvin Probe Force Microscopy is a method to assess the contact potential difference between a sample and the probe tip. It remains a relative tool unless a reference standard with a known work function is applied, typically bulk gold or cleaved highly oriented pyrolytic graphite. In this report, we suggest a verifiable, two-dimensional standard in the form of a photolithographically patterned, wire-bonded structure manufactured in the technology of transfer-free p-type hydrogen-intercalated quasi-free-standing epitaxial Chemical Vapor Deposition graphene on semi-insulating high-purity nominally on-axis 4H-SiC(0001). The particular structure has its hole density pS= 1.61 × 1013 cm−2 measured through a classical Hall effect, its number of the graphene layers N= 1.74 extracted from the distribution of the ellipsometric angle Ψ, measured at the angle of incidence AOI = 50° and the wavelength λ= 490 nm, and its work function ϕGR= 4.79 eV postulated by a Density Functional Theory model for the specific pS and N. Following the algorithm, the contact potential difference between the structure and a silicon tip, verified at ΔVGR−Si=0.64V, ought to be associated with ϕGR= 4.79 eV and applied as a precise reference value to calculate the work function of an arbitrary material.
No takes yet. Share an insight, caveat, or question.
Ciuk et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: