Chronic kidney disease (CKD) is a progressive condition that affects more than 10% of the world's population. Monitoring urine albumin-to-creatinine ratio (uACR) has become the gold standard for nephropathy diagnosis and control. The objective of the present study was to develop a simple, accurate, sensitive, and rapid point-of-care test (PoCT) device, MyACR, for uACR measurement, intended for use in community healthcare to screen for the risk and monitor the progress of CKD. Albumin and creatinine concentrations in urine samples were determined using spectrophotometric dye (tetrabromophenol blue)-binding and colorimetric Jaffe assay, respectively. Urine samples were diluted with distilled water (1:80) and mixed separately with albumin and creatinine reaction mixture. The creatinine reaction was incubated at room temperature (25 °C) for 30 min before analysis. Optical density (OD) was measured at the wavelengths of 625 nm (albumin) and 515 nm (creatinine). All calibration curves (0-60 mg/L and 0-2 mg/dL for albumin and creatinine) yielded linear relationships with correlation coefficients (R