State-of-the-art optical cavities are pivotal in pushing the envelope of laser frequency stability below 10−16. This is often achieved by extending the cavity length or cooling the system to cryogenic temperatures to reduce the thermal noise floor. In our study, we present a 30-cm-long cavity that operates at room temperature and is outfitted with crystalline coatings. The system has a predicted ultralow thermal noise floor of 4.4 × 10−17, comparable to what is observed in cryogenic silicon cavities. A 1397-nm laser is stabilized in this advanced cavity, and the stable frequency is then transferred to the clock transition in strontium optical lattice clocks via a frequency-doubling process. We have meticulously minimized and assessed the technical noise contributions through comparisons with an ultrastable reference laser that is locked to a commercially available 30-cm cavity. The frequency instability of the system is rigorously evaluated using a three-cornered-hat method. The results demonstrate that the laser frequency instability remains below 2 × 10−16 for averaging times ranging from 1 to 50 s. These findings underscore the significant potential of room-temperature cavities with crystalline coatings in high-precision metrology and pave the way for further improvements in optical lattice clocks.