To address the inadequacies of traditional seawater total alkalinity detection methods, which fail to meet the demands for continuous on-site monitoring at sea, a seawater total alkalinity analyzer based on single-point titration and spectrophotometry has been developed. This paper focuses on the design of the analyzer’s system architecture, hardware, and control systems. The instrument adopts a modular design concept and utilizes microfluidic technology for detection, with the STM32 serving as the control core. Through the design of the control system for the detection instrument, accurate in-situ automatic measurement of seawater total alkalinity is achieved. Laboratory testing and analysis demonstrate that the instrument’s measurement relative error is less than 1%, satisfying the accuracy requirements for seawater total alkalinity measurement. This research provides critical monitoring data support for a timely understanding of marine ecological environments and the marine carbon cycle.