Real-time and accurate temperature monitoring during microwave hyperthermia (MH) remains a critical challenge for ensuring treatment efficacy and patient safety. This study presents a novel approach to simulate real MH and precisely determine the temperature of the target region within biological tissues using a temporal-informed neural network. We conducted MH experiments on 30 sets of phantoms and 10 sets of ex vivo pork tissues. We proposed a novel perspective: the evolving tissue responses to continuous electromagnetic radiation stimulation are a joint evolution in temporal and spatial dimensions. Our model leverages TimesNet to extract periodic features and Cloblock to capture global information relevance in two-dimensional periodic vectors from ultrasound images. By assimilating more ultrasound temporal data, our model improves temperature-estimation accuracy. In the temperature range 25-65 °C, our neural network achieved temperature-estimation root mean squared errors of approximately 0.886 °C and 0.419 °C for fresh ex vivo pork tissue and phantoms, respectively. The proposed temporal-informed neural network has a modest parameter count, rendering it suitable for deployment on ultrasound mobile devices. Furthermore, it achieves temperature accuracy close to that prescribed by clinical standards, making it effective for non-destructive temperature monitoring during MH of biological tissues.