Titanium plates are the current gold standard for fracture fixation of the mandible. Magnesium alloys such as WE43 are suitable biodegradable alternatives due to their high biocompatibility and elasticity modulus close to those of cortical bone. By surface modification, the reagibility of magnesium and thus hydrogen gas accumulation per time are further reduced, bringing plate fixation with magnesium closer to clinical application. This study aimed to compare bone healing in a monocortical mandibular fracture model in sheep with a human-standard size, magnesium-based, plasma electrolytic-oxidation (PEO) surface modified miniplate fixation system following 4 and 12 weeks. Bone healing was analyzed using micro-computed tomography and histological analysis with Movat's pentachrome and Giemsa staining. For evaluation of the tissue's osteogenic activity, polychrome fluorescent labeling was performed, and vascularization was analyzed using immunohistochemical staining for alpha-smooth muscle actin. Bone density and bone mineralization did not differ significantly between titanium and magnesium (BV/TV: T1: 8.74 ± 2.30%, M1: 6.83 ± 2.89%,
No takes yet. Share an insight, caveat, or question.
Turostowski et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: