Positron Emission Tomography (PET) stands out as a highly specific molecular imaging technique. However, its detection sensitivity remains a challenge. The implementation of Time-of-Flight (TOF) PET technology enhances sensitivity by precisely measuring the time lapse between the annihilation photons. Moreover, by characterizing scattered (Compton) events, the effective sensitivity of PET imaging might significantly be enhanced. In this work, we present the scatter sub-system of a 2 layers preclinical TOF-PET scanner for mice head imaging. The scatter sub-system is composed of 8 identical modules based on analog SiPMs coupled to crystal arrays of 24 × 24 LYSO pixels with 0.95 mm × 0.95 mm × 3 mm dimensions. The system has 29 mm bore and 50.8 mm axial length. An average CTR of 192± 1 ps was obtained for the whole sub-system at the photopeak energy range after energy and timing corrections, and CTR values as good as 155 ps were found for some individual pixels. The transit time spread at the SiPM level was also studied and corrected, achieving a mean value of 41 ps of maximum time difference at the sensor corners with respect to the center. Voronoi diagrams were implemented to correct for position decoding.