Performance evaluation shows high spatial resolution and sensitivity in small-animal PET imaging, indicating advanced capabilities for preclinical studies.
This work reports the performance evaluation of a newly developed small-animal positron emission tomography (PET) system based on lutetium-yttrium oxyorthosilicate (LYSO) crystals and multi-pixel photon counter (MPPC). Performance was evaluated, including spatial resolution, system sensitivity, energy resolution, scatter fraction (SF), noise–equivalent count rate (NECR), micro-Derenzo phantom imaging, and in vivo imaging of mice and rats. The system achieved a tangential spatial resolution of 0.9 mm in the axial direction at a quarter axial offset using the three-dimensional ordered-subsets expectation maximization (3D OSEM) reconstruction algorithm. The peak sensitivity was 8.74% within a 200–750 keV energy window, with an average energy resolution of 12.5%. Scatter fractions were 12.9% and 30.0% for mouse- and rat-like phantoms, respectively. The NECR reached 878.7 kcps at 57.6 MBq for the mouse phantom and 421.4 kcps at 63.2 MBq for the rat phantom. High-resolution phantom and in vivo images confirmed the system’s capability for quantitative, high-sensitivity small-animal imaging, demonstrating its potential for preclinical molecular imaging studies.