The early detection and treatment of Helicobacter pylori infections can help prevent the formation of ulcers and development of gastric cancer in humans. The urea breath test (UBT) is a noninvasive test for diagnosing H. pylori . However, isotopic‐labeled UBT reagents are expensive and require specialized test equipment to detect isotopic CO 2 . In this study, the development of a user‐friendly, cost‐effective, and alternative method for diagnosing H. pylori infections by detecting NH 3 in exhaled breath is reported. A commercial disposable pH strip is modified with sodium carboxymethylcellulose to enhance its performance in gaseous environments. The modified pH strip exhibits compatibility with both acidic (HCl and CH 3 COOH) and alkaline (NH 3 ) gases. Furthermore, the sensor monitors the production of NH 3 resulting from fish spoilage for 72 h, highlighting its real‐time capabilities. The feasibility of employing this simple strip sensor to detect H. pylori infection in patients is investigated through in vitro experiments and preliminary clinical studies. The simple strip sensor is a promising alternative to the C‐13 UBT for detecting H. pylori infection.
No takes yet. Share an insight, caveat, or question.
Tsai et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: