Emerging point-of-care testing methods are extremely beneficial for personalized assessments of trace element metabolism including selenium (Se). Given the lack of timely evaluation methods for well-received Se fortification, an electrochemical solution was developed based on the recently identified urinary selenosugar (Sel) as a marker. The Se content of crude urine was rapidly determined (∼5 min), and the square-wave voltammetric responses of a Se-selective probe (SeSE) composed of liquid metal amalgam demonstrated comparable performance (e.g., detection limit: 19 nM) to central lab benchtop equipment within the physiological range. Meanwhile, SeSE enabled total urinary Se detection via a mere one-step oxidation. Additionally, SeSE was utilized to jointly assess the apparent internalization and utilization rate of two typical nutrients, selenite and selenomethionine, in a rat nutrition model, demonstrating consistent results with those obtained by HPLC-MS and ICP-MS. Upon systematic standardization directed by Ramaley's theory, SeSE was integrated into a battery-operated portable kit (dubbed "SeEye") with a micro electrochemical drive and tablet PC console for one-stop service trials in a local commercial scenario. This study establishes (1) a nutritive value classifier in a low-cost consumer electronic format and (2) noninvasive diagnostic technology for Se supplementation.
No takes yet. Share an insight, caveat, or question.
Song et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: