In this study, we present the development of a loop-based two-dimensional supercritical fluid system in multiple heart-cutting modes (mSFC-SFC), with diode-array and mass spectrometric detection. The instrument design was developed to be as simple as possible, based on a single SFC instrument, with the sole addition of three external 2-port 6-position valves. The objective was to achieve the most complete transfer of a peak from the first to the second dimension, whatever the composition of the mobile phase, i.e., whatever the proportion of carbon dioxide and methanol cosolvent along a wide gradient elution. Thanks to fine adjustment of the valve switching times, the first-dimension peaks were parked in 50 μL or 100 μL loops and later discharged to the second dimension. The interest of this instrument was then demonstrated with a sample application on a natural product: an extract of